The failed module in the detector has been replaced. In the process of this repair, we have noticed some issues with a chip in a module at the bottom of the detector. A Dectris engineer will be on-site Dec 2 to address this issue, which they believe is due to a loose cable.
The following command can be used in autoPROC to mask out the bad chip:
autoPROC_XdsKeyword_UNTRUSTED_RECTANGLE="798 859 2429 2526"
or use the following line in your XDS.INP file:
UNTRUSTED_RECTANGLE=798 859 2429 2526